Hantek — :ACQuire:AVERages

:RUN :STOP For example, :RUN /*The oscilloscope starts to run*/ :STOP /*The oscilloscope stops running*/ :SINGle For example, :SINGle / *Single acquisition of oscilloscope*/ :TFORce For example, :TFORce / *Forced oscilloscope triggering*/ :ACQuire Command Subsystem The :ACQuire command is used to set and query the oscilloscope's storage depth, acquisition method and average number of samples, and query the current sampling rate. :ACQuire:AVERages 16 / 136 Command format :RUN :STOP Function description : RUN starts the oscilloscope. : STOP command stops the oscilloscope. This command is equivalent to pressing the RUN/STOP button on the front panel.

Description This command is invalid when the waveform recording function is turned on or when the recorded waveform is played back. Command format :SINGle Function description Set the oscilloscope to single trigger mode. This command is equivalent to pressing the SINGLE button on the front panel. Description ▷In single trigger mode, the oscilloscope will trigger once when the trigger conditions are met, and then stop. ▷This command is invalid when the waveform recording function is turned on or when the recorded waveform is played back. Command format :TFORce Function description Force a trigger signal. Suitable for normal and single trigger mode.

This command has the same function as pressing the front panel trigger control area For ForceTrig key. Command format :ACQuire:AVERages :ACQuire:AVERages? Function description Set or query the average number of times in the average acquisition mode Parameter Name Type Range Default integer 2𝑛(n is an integer, ranging from 1 to 10) 2