Hantek — :CURSor:MANual:VUNit0

For example, :CURSor:MANual:TYPE X /*Select X type cursor*/ :CURSor:MANual:TYPE? /*The query returns X*/ :CURSor:MANual:SOURce For example, :CURSor:MANual:SOURce CHANnel2 /*Set the channel source to CH2*/ :CURSor:MANual:SOURce? /*The query returns CHAN2*/ :CURSor:MANual:TUNit0 For example, :CURSor:MANual:TUNit DEGRee /*Set the horizontal unit to "degree"*/ :CURSor:MANual:TUNit? /*The query returns DEGR*/ :CURSor:MANual:VUNit0 24 / 136 parameters. Return format The query returns X or Y. Command format :CURSor:MANual:SOURce :CURSor:MANual:SOURce? Function description Set or query the channel source for manual cursor measurement.

Parameter Name Type Range Default Discrete {CHANnel1|CHANnel2|CHANnel3| CHANnel4|MATH|LA} CHANnel1 Description ▷Only the currently opened channel can be selected as the channel source. ▷When LA is selected, the cursor type cannot be set to Y (: CURSor: MANual: TYPE). Return format The query returns CHAN1, CHAN2, CHAN3, CHAN4, MATH, or LA. Command format :CURSor:MANual:TUNit :CURSor:MANual:TUNit? Function description Sets or queries the horizontal unit in the manual cursor measurement mode. Parameter Name Type Range Default Discrete {S|HZ|DEGRee|PERCent} S Description ▷S: AX, BX, and BX-AX in the measurement results are in seconds, and 1 / | dX | is in hertz. ▷HZ:

AX, BX, and BX-AX in the measurement result are in "Hertz", 1 / | dX | is in "Seconds". ▷DEGRee: AX, BX and BX-AX in the measurement results are in "degrees". ▷PERCent: AX, BX and BX-AX in the measurement results are expressed in percentage format. Return format The query returns S, HZ, DEGR, or PERC. Command format :CURSor:MANual:VUNit :CURSor:MANual:VUNit? Function description Sets or queries the vertical unit in manual cursor measurement mode. Parameter Name Type Range Default Discrete {PERCent|SOURce} SOURce