Hantek — :MEASure:SOURce

For example, :MASK:Y 0.36 /*Set the vertical adjustment parameter to 0.36div*/ :MASK:Y? /*The query returns 3.600000e-01*/ :MASK:CREate :MASK:PASSed? :MASK:FAILed? :MASK:TOTal? :MASK:RESet :MEASure Command subsystem :MEASure:SOURce 43 / 136 Function description Sets or queries the vertical adjustment parameter in the pass / fail test rule. The default unit is div. Parameter Name Type Range Default Integer 0.04 to 5.12 in 0.04 steps 0.96 Return format The query returns the current vertical adjustment parameter in scientific notation. Command format :MASK:CREate Function description Create a pass / fail rule with the currently set horizontal and vertical adjustment parameters.

Description This command is valid only when the pass / fail test function is turned on (: MASK: ENABle) and is not running (: MASK: OPERate). Command format :MASK:PASSed? Function description Query the number of frames passed in the pass / fail test. Return format The query returns an integer. Command format :MASK:FAILed? Function description Query the number of frames that failed in the pass / fail test. Return format The query returns an integer. Command format :MASK:TOTal? Function description Query the total number of frames that passed / failed the test. Return format The query returns an integer.

Command format :MASK:RESet Function description Resets the number of frames passed, failed frames, and total frames in the pass / fail test. Command format :MEASure:SOURce :MEASure:SOURce? Function description Set or query the source of the current measurement parameters. Parameter Name Type Range Default