For example, :TRIGger:DELay: WHEN GOUT /*Set delay type to> :TRIGger:DELay: WHEN? /*The query returns GOUT*/ :TRIGger:DELay:TIME For example, :TRIGger:INTERVAl:TIME 0.000003 /*Set the pulse width value 3us*/ :TRIGger:INTERVAl:TIME? /*The query returns 3.000000e-06*/ :TRIGger:SHOLd :TRIGger:SHOLd:DSrc For example, :TRIGger:SHOLd: DSrc CHANnel1 /*Set the data source to CH1*/ :TRIGger:SHOLd: DSrc? /*The query returns CHAN1*/ :TRIGger:SHOLd:CSrc 84 / 136 Command format :TRIGger:DELAY:TIME :TRIGger:DELAY:TIME?
Function description Set or query the time value when the delay is triggered Parameter Name Type Range Default Discrete 8ns-10s 20ns Return format Returns the time value in scientific notation Command format :TRIGger:SHOLd:DSrc :TRIGger:SHOLd:DSrc? Function description Sets or queries the data source for establishing a hold trigger. Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22| D23|D30|D31|D32|D33|D40|D41| D42|D43| CHANnel1|CHANnel2|CHANnel3| CHANnel4} CHANnel1 Description When the digital channel is used as the trigger source channel, the digital probe must be connected to the oscilloscope.
Return format The query returns D10, D11, D12, D13, D20, D21, D22, D23, D30, D31, D32, D33, D40, D41, D42, D43, CHAN1, CHAN2, CHAN3, CHAN4. Command format :TRIGger:SHOLd:CSrc :TRIGger:SHOLd:CSrc? Function description Set or query the clock source for establishing hold trigger. Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22 |D23|D30|D31|D32|D33|D40|D41| D42|D43|CHANnel1|CHANnel2| CHANnel2