Basic Operation The Autoset function examines all channels for signals and displays corresponding waveforms. Autoset determines the trigger source according to the following conditions. ➢ If multiply channels get signals, the oscilloscope will use the channel with the lowest frequency signal as the trigger source. ➢ If no signals are found, the oscilloscope will use the lowest-numbered channel displayed in Autoset as the trigger source. ➢ If no signals are found and no channels are displayed, the oscilloscope will display and use Channel 1 as the trigger source. Sine Wave:

When you use the Autoset function and the oscilloscope determines that the signal is similar to a sine wave, the oscilloscope displays the following options. Square Wave or Pulse: When you use the Autoset function and the oscilloscope determines that the signal is similar to a square wave or pulse, the oscilloscope displays the following options.

DSO1000B(V) Series HandHeld Oscilloscope User Manual 38 TIME/DIV Adjusted Trigger Coupling Adjusted to DC, Noise Reject or HF Reject Trigger Holdoff Minimum Trigger Level Set to 50% Trigger Mode Auto Trigger Source Adjusted Trigger Slope Adjusted Trigger Type Edge Trigger Video Sync Adjusted Trigger Video Standard Adjusted Vertical Bandwidth Full Vertical Coupling DC (if GND was chosen before); AC for the video signal; otherwise, unchanged VOLTS Adjusted Sine Wave Options Details Multi-cycle Sine Display multiple cycles that have appropriate vertical and horizontal scales. Single-cycle Sine Set the horizontal scale to display about one cycle of the waveform.

Cancel Let the oscilloscope recall the previous setup. Square Wave Options Details Multi-cycle Display multiple cycles that have appropriate vertical and horizontal scales. Single-cycle Set the horizontal scale to display about one cycle of the waveform. The oscilloscope displays Min, Mean and Positive Width automatic measurements. Rising Display the rising edge. Falling Display the falling edge. Cancel Let the oscilloscope recall the previous setup.