Hantek — 5.2 Accuracy

Chapter 5 Specification The general and measurement precision specification of LCR are described below. 5.1 General Specification 5.2 Accuracy Notes:  Environment temperature: 20 ℃ ± 2 ℃ , humidity: ≤75% R.H;  Preheat the instrument for at least 30 minutes before the test;  Test at the test notch on the end face of the instrument;  It is better to conduct open and short circuit correction before the test;  Measure with the recommended equivalent mode;  The percentage in terms of error indicates:

 ± (% of the reading + last digit)  If the actual measurement of the instrument and the display exceeds the scope specified in the table, the accuracy of the excessive part will not be given;  The meaning of the subscript  S- series equivalent; p- parallel equivalent; e- accuracy  Some parameters cannot be given in the data table, and it can only be calculated based on the measurement results; 23 Model Hantek1832C Hantek1833C Measurement parameters Main parameter: L/C/R/Z Secondary parameter:

X/D/Q/θ/ESR Equivalent mode Series, parallel Mode of range Manual, automatic Measurement speed Fast(4 times/s), medium(2 times/s), slow(1 times/s) Configuration of the test terminal 3-terminal, 5-terminal Correction function Short circuit, Open circuit Communication interface Type C(virtual serial port) Test signal frequency 100Hz, 120Hz, 400Hz, 1kHz, 4KHz, 10kHz, 40kHz 100Hz, 120Hz, 400Hz, 1kHz, 4KHz, 10kHz, 40kHz, 50KHz, 75KHz, 100kHz Test signal level 0.6Vrms 0.3Vrms, 0.6Vrms Output impedance 100Ω Highest accuracy (see accuracy index for details) Resistance: 0.25% Capacitance: 0.4% Measurement range L: 0-2000H C: 0-20mF R: 0-20MΩ