Hantek — The front panel

the [Utility] function: The front panel menu references The following table summarizes the front panel keys and menu structure. 11 Keys Application Configure DC voltage measurement, including DCV ratio measurement: Range: Automatically adjust range (default), 100 mV, 1 V, 10 V, 100 V or 1000V Aperture PLC: 0.02, 0.2, 1, 10, 100. Default value: 10 Auto Zero: Off or On (default) Input Z: 10 M Ω (default) or Auto Ω (> 1 G) DCV ratio: Off (default) or On Configure DC current measurement: Terminals: 3 A or 10 A Range: Auto, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A or 10 A (Terminals set to 10 A). Aperture PLC: 0.02, 0.2, 1, 10, 100. Default value: 10 Auto Zero:

Off or On (default) Configure AC voltage measurement: Range: Automatically adjust range (default), 100 mV, 1 V, 10 V, 100 V or 750 V AC Filter: >3 Hz, >20 Hz, >200 Hz Configure AC current measurement: Terminals: 3 A or 10 A Range: Auto, 100 µA, 1 mA, 10 mA, 100 mA, 1 A, 3 A or 10 A(terminal set to 10 A) AC Filter: >3 Hz, >20 Hz, >200 Hz