Downloaded from https://www.everyspec.com AMCP 706-117 where TABLE 7-7. APPARENT DETECTOR TIME Y = instantaneous value of the CONSTANT VS ME-...R READING microphone output voltage T = interval of time over which the Time constant c, Deviatlon AL, integration is carried out ns dB The rms detector must, therefore, include a squaring circuit, an integrator, and a root extraction network. 53 - 0.1 Since many sound signals are nonsinusoidal, the detec- 100 - 0.63 tor must be able to handle a relatively high crest factor (ratio of peak to rms voltage).

Many circuits have been 127 - I developed to approximate the square-law function, but ptrhaps the most widely used circuit is the quasi.rms 200 - 2 detector. In this detector an integrating capacitor 390 - 4 biases a rectifier so that no current flows until the absolute value of the input signal exceeds the voltage on the capacitor. Above this point, a linear relat: nship In response to a single 0.b-s burst of a 1-kHz tone, the exists between the input signal and the detector cur- sound4evel meter should indicate a nominal 4 dB below rent. By appropriate selection of component values, the steady-state value.

IEC 179 puts a ± 1 dB tolerance deviation from the ideal is held within acceptable limits on the reading; IEC 123 and 81.4, ± 2 dB. The "slow" (usually ± I dB) up to a crest factor of three. The mode reduces meter jitter In measurements of rapidly quasi.tins detector more than satisfies the requirements varying sound fields by providing a longer averaging of IEC and USA standards for sound-level meters. In time. For "slow", the nominal detector time constant Is practice, the detector i6 a Pesistance-capacitance about 1 s. (RC) averaging detector wherein the RCtime constant (3) Measurements with the sound level meter.

In the is chosen to provide meter response appropriate to that final analysis, the sound-level meter measures the fre- stipulated in the standards for "fast" and "slow" detec- quency- and time-weighted rms value of sound pres- tion characteristics. The standards stipulate overall sure. The frequency weighting is selected Ls dB(A), sound-level meter performance (designated "fast" in dB(B), or dB(C). As noted previously, thes response the IEC standards) in response to a single 200-ms burst curves make the sound-level meter respond to single of a l-kHz sinusoidal signal.

IEC 179 requires the meter tones at various frequencies in approximately the same to read I dB ± I dB below the steady-state value, while way as the human ear. As broadband devicms, however, IEC 123 and S.4 require a meter reading between 0 and sound-level meters cannot indicate the spectral compo- 4 dB below the steady-state value. sition of a particular sound; i.e., no correlation exists The reduction in meter rea,.ing (from the steady- between our subjective response due to this factor and state value) for various apparent RCtime constarnts can meter indication.

This is the main reason that substan- be calculated from the formula tial differences--10 dB and more-can exist betwcen subjectively measured loudness level in phons and & = 10 log [1 - eXp (- tic)] (77) sound-pressure level in dB. Furthennore, other factors that contribute include the inability of the sound-level meter to account for masking effects (discussed in par. •,here 7-5) and the impossibility of selecting the right weight- AL = meter reading in dB below the ing function for all spectral components at once.

steady-statu value Since no reasonable degree of correlation between I - duration of the tone burst, 200 the loudness levels of widely differing sounds and ms in this case sound-evel meter readings can be achieved, usually no c = apparent time constant of the attempt is made to select the most appropriate weight- detector :ng curve-A, B, or C-for the levels encountered.

Thre (The formula is valid provided that the crest factor quantity that is measured is almost always the A- capability of the detector is not exceeded, and instru- weighted sound level, even at ivvels where the high ment response is determined by the detector.) Table 7-7 attenuation of low frequency components would not be shows some values of apparent time constants and cor- justified from a physiological point of view. responding meter readings. 7-16