For example, :TRIGger:CAN:DLC 10 /*Set the data length code to 10*/ :TRIGger:CAN:DLC? /*The query returns 10*/ :TRIGger:CAN:DATA For example, :TRIGger:CAN:DATA 2 10 /*Set the data with index 2 to 10*/ :TRIGger:CAN:DATA? 2 /*The query returns 10*/ :TRIGger:CAN:ALEVel For example, :TRIGger:CAN:ALEVel 0.16 /*Set trigger level 160mV*/ :TRIGger:CAN:ALEVel? /*The query returns 1.600000e-01*/ :TRIGger:IIC :TRIGger:IIC:SDA:SOURce 96 / 136 Return format The query returns an integer Command format :TRIGger:CAN:DATA :TRIGger:CAN:DATA?
Function description Set or query CAN trigger data value Parameter Name Type Range Default Discrete 0-3 Integer 0~255【XX】 XX Return format The query returns an integer Command format : TRIGger:CAN:ALEVel : TRIGger:CAN:ALEVel? Function description : Set or query the trigger level during CAN trigger Parameter Name Type Range Default Integer (-5 × VerticalScale - OFFSet) to (5 × VerticalScale - OFFSet) 0 Description This setting command is valid only when the channel source of the data line is an analog channel. Return format Returns the trigger level value in scientific notation Command format :TRIGger:IIC:SDA :TRIGger:IIC:SDA?
Function description Set or query the channel source of the I2C triggered data line. Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22 |D23|D30|D31|D32|D33|D40|D41 |D42|D43|CHANnel1|CHANnel2| CHANnel3|CHANnel4} CHANnel2 Description When the digital channel is used as the trigger source channel, the digital probe must be connected to the oscilloscope.