:TRIGger:IIC:AWIDth For example, :TRIGger:IIC:AWIDth 10 /* Set the address bit width to 10*/ :TRIGger:IIC:AWIDth? /* The query returns 10*/ :TRIGger:IIC:CLEVel For example, :TRIGger:IIC:CLEVel 0.16 /* Set the trigger level to 160mV */ :TRIGger:IIC:CLEVel? /* The query returns 1.600000e-01*/ :TRIGger:IIC:DLEVel For example, :TRIGger:IIC:DLEVel 0.16 /* Set the trigger level to 160mV*/ :TRIGger:IIC:DLEVel? /* The query returns 1.600000e-01*/ 99 / 136 Command format :TRIGger:IIC:AWIDth :TRIGger:IIC:AWIDth? Function description Set or query the address bit width when the trigger condition of I2C is address or address data.
Parameter Name Type Range Default Discrete {7|10} 7 Return format The query returns 7, 8, or 10. Command format :TRIGger:IIC:CLEVel :TRIGger:IIC:CLEVel? Function description Set or query the trigger level of the clock line when I2C triggers. The unit is consistent with the current amplitude unit. Parameter Name Type Range Default Real (-5 × VerticalScale - OFFSet) to (5 × VerticalScale - OFFSet) 0 Description This setting command is valid only when the channel source of the clock line is an analog channel. Return format The query returns the trigger level in scientific notation. Command format :TRIGger:IIC:DLEVel :TRIGger:IIC:DLEVel?
Function description Set or query the trigger level of the data line when I2C triggers, and the unit is consistent with the current amplitude unit. Parameter Name Type Range Default Real (-5 × VerticalScale - OFFSet) to (5 × VerticalScale - OFFSet) 0 Description For VerticalScale, please refer to the :CHANnel :SCALe command, and for OFFSet, please refer to the :CHANnel :OFFSet command. This setting command is valid only when the channel source of the data line is an analog channel. Return format The query returns the trigger level in scientific notation.