User Manual 1. Press the [Trig Menu] button on the front panel to enter the Trigger function menu. 2. Press the Type softkey, turn the Multifunctional Knob to set select Slop and then press the knob to confirm. 3. Press the Source softkey, turn the Multifunctional Knob to select CH1~CH2 as the trigger source. 4. Press the Slop softkey, turn the Multifunctional Knob to set select the desired trigger edge (rising or falling), and then push down the knob to confirm. 5. Press the Level softkey and press of Lower Upper softkey to select Lower(V2) or Upper(V1) the trigger level; then turn the Trigger Level Knob to adjust the position.
The trigger different level value between Upper and Lower is displayed at the upper right corner of the screen. The Lower trigger level cannot be upper than the upper trigger level. V1 means the upper trigger level while V2 means the lower trigger level. 6. Press the When softkey, then turn the Multifunctional Knob to select the desired slope condition, and push down the knob to confirm. ( less than a time value ): trigger when the positive or negative slope time of the input signal is lower than the specified time value. > ( greater than a time value ): trigger when the positive or negative slope time of the input signal is greater than the specified time value.
!=( not equal to a time value ): trigger when the positive or negative slope time of the input signal is not equal to the specified time value.. =( equal to a time value ): trigger when the positive or negative slope time of the input signal is equal to the specified time value. 7. Press the Time softkey and turn V0 to set the reference value of the slope time. 8. Press the Mode softkey, turn V0 to select the trigger mode (auto, normal), and press V0 to confirm. Auto : When the oscilloscope meets the trigger condition, it completes a trigger acquisition once; when the trigger condition is not met, it can run the acquisition waveform freely. Normal :
When the oscilloscope meets the trigger condition, the input waveform is displayed; when the trigger condition is not met, the original waveform is displayed. 9. Press the Holdoff softkey and turn V0 to set the time that the oscilloscope waits before a trigger to the next trigger, so that complex waveforms are displayed stably. 2.7.5. Overtime Trigger Trigger when the time interval ( △ T) from when the rising edge (or falling edge) of the input signal passes through the trigger 24 / 67