Hantek — 2.10.3.Automatic Measurement

User Manual To Make All Measurement All measurement could measure all parameters of the current measurement source and display the results on the screen. Do the following steps to make all parameters measurement. 1. Press the [Meas] button on the front panel to enter the MEASURE function menu. 2. Press the All Measure softkey to select ON. 3. Press the Source softkey to select the measure source (CH1~CH2). 1 Frequency The reciprocal of the period. 2 Period The time between two consecutive threshold points of the same polarity edge. 3 Average The arithmetic mean of the entire waveform or selected area. 4 Pk-Pk The voltage value from the peak to the lowest point of the waveform.

5 RMS That is a valid value. According to the energy converted by the AC signal in one cycle, the DC 6 Period Rms The root mean square value of the signal within 1 cycle. voltage corresponding to the equivalent energy is the root mean square value. 7 Min The most negative peak voltage measured over the entire waveform. 8 Max The most positive peak voltage measured over the entire waveform. 9 RiseTime Measure the time between 10% and 90% of the first rising edge of the waveform. 10 FallTime Measure the time between 90% and 10% of the first falling edge of the waveform. 11 + Width Measure the time between the first rising edge and the next falling edge at the waveform 50% level.

Measure the time between the first falling edge and the next rising edge at the waveform 50% 12 - Width level. 13 + Duty Measure the first cycle waveform. Positive Duty Cycle is the ratio between positive pulse 45 / 67 No. Type Comments