Hantek 6000EU Digital Oscilloscope ■ From the “Probe” menu, select attenuation to 1:10. Set the switch to “X10” on the probe and connect it to CH1 of the oscilloscope. When using the probe hook-tip, insert the tip onto the probe firmly to ensure a proper connection. ■ Attach the probe tip to the Probe Compensator and the reference lead to the ground connector, select CH1, and then press the “ AUTOSET“ button into the menu or the toolbar. ■ Check the shape of the displayed waveform. Correctly Compensated Over Compensated Under Compensated 1. I f necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the fattest square wave being displayed on the oscilloscope. 2.