Hantek 6000EU Digital Oscilloscope 3. Attach the tip of probe and ground nip to the Connector of Probe compensator. Click the button. A square wave will be displayed within a several seconds. (Approximately 1 kHz, 2V, peak- to- peak). 4. Inspect CH2 with the same method. Repeat steps 2 and 3. 1.6 Self Calibration The self calibration routine lets you optimize the oscilloscope signal path for maximum measurement accuracy. You can run the routine at any time but you should always run the routine if the ambient temperature changes by 5 ℃ or more. For accurate calibration, power on the oscilloscope and wait twenty minutes to ensure it is warmed up.
To compensate the signal path, disconnect any probes or cables from the input connectors. Then, access the “Utility -> Calibratio n” option and follow the directions on the screen. The self calibration routine takes about several minutes. USER’S MANUAL 14