Application measurement EN Figure 6-17 System settings 6.5 Calibration Click on [Utility] -> Cal in sequence, enter the password, and you can set [Zero], [Calib Vol], and [Calib Res]. The specific steps are as follows: Click Cal to pop up the numeric keypad for the password. After entering the correct password, enter the calibration submenu; If the password is incorrect, a prompt message "Password Error" will pop up on the interface. Zero Details reference Appendix C:
Performing Zero Adjustment Voltage calibration setting steps 1 Connect the DC power supply to the input port of the internal resistance tester and set the power output 2 Input actual voltage Click on Calib Vol -> Real Vol in sequence, and the page will display Voltage Calibration (P). Use the left/right keys to move the cursor, click [OK] to enter the actual voltage value, position the cursor to the key , and then click [OK] to save the value. Copyright © Qingdao Hantek Electronic Co., LTD HBT3000 series user manual 37