Modulation EN 8.13.5 Modulation Frequency When internal modulation source is selected, press [Mod] > Frequency softkey to set the frequency of the modulating waveform. ⚫ Use the numeric keypad or direction keys and knob to enter the desired frequency value. ⚫ The modulating waveform frequency range is 2mHz to 1MHz, and the default value is 100Hz. Note: This menu will be grayed out and disabled when other modulation sources except the internal modulation source is selected. 8.13.6 Duty Deviation Press [Mod] > Deviation softkey and input the desired value using the numeric keypad or direction keys and knob.
Duty deviation represents the variation (in %) of the modulated waveform Duty relative to the original pulse Duty. ⚫ Duty deviation range: 0.1% to 49.9%. ⚫ The Duty deviation cannot exceed the current pulse Duty. ⚫ Duty deviation is limited by the minimum Duty and current edge time setting. When external modulation source is select, the Duty deviation is controlled by the ±4V signal level on the [FSK/Trig/Sync/Extmod] connector on the rear panel Copyright © Qingdao Hantek Electronics Co., LTD HDG3000B User manual 57