Command system EN FUNCtion:PULSe:DPeriod and FUNCtion:PULSe:WIDTh commands affect the same arguments. ⚫ Pulse width is the time from the 50% threshold on the rising edge to the 50% ⚫ Pulse width ≥ minimum pulse width threshold on the next falling edge. Pulse width ≤ pulse period - 2 x minimum pulse width Return format Return the pulse width value as a scientific count. For example, SOURce1: FUNCtion: PULSe, WIDTh of 0.005 /* Set channel 1 pulse width to 5ms */ 3.13.9 SOURce :FUNCtion:PULSe:TRANsition:LEADing Command format SOURce :FUNCtion:PULSe:TRANsition:LEADing |MINimum|MAXimum SOURce :PULSe:TRANsition:LEADing? [MINimum|MAXimum] Function description Set pulse rising edge time.
Argument None Instructions ⚫ The leading edge time applies to the rising edge and represents the time between the 10% threshold and the 90% threshold of the edge. ⚫ The specified edge time must be within the specified pulse width and period. The instrument will limit the edge time to meet the specified pulse width or duty cycle. Return format Return the time value as a scientific count. Example SOURce1: FUNCtion: PULSe: the TRANsition: LEADing 0.00000001 /* Set Channel 1 pulse rising time to 10ns */ 3.13.10 SOURce :FUNCtion:PULSe:TRANsition:TRAiling Command format HDG300C Programming manual Copyright © Qingdao Hantek Electronics Co., LTD 40