Hantek — Query Current Measurement Value:MEASure:CURRent?

Channel 3: 0V-8.1V 4 channels model: Channel 1: 0V-32.1V Channel 2: 0V-32.1V Channel 3: 0V-8.1V Channel 4: 0V-16.1V Return value The query returns the overvoltage protection voltage setting value of the query channel, unit V. For example VOLTage:PROTection 30.5,(@2) /* Set the overvoltage protection voltage of channel 2 to 30.5V */ VOLTage:PROTection? (@2) /*Return 30.5*/ MEASure Subsystem Query Voltage Measurement Value : MEASure:VOLTage? Command format MEASure:VOLTage? Function description Query the measured value of the channel voltage. Parameter ch: Refer to OUTP command. Description None. Return value The query returns the measured voltage value of the channel, unit V.

For example MEASure:VOLTage? (@1) Query Current Measurement Value : MEASure:CURRent? Command format MEASure:CURRent? Function description Query channel current measurement value. Parameter 11