Hantek — 5.5 Probe compensation

Preparation before use EN The default language is English. 5.4 Functionality check Click [OSC] to enter the oscilloscope interface. Click Utility > Default in the lower left corner of the screen to restore the oscilloscope to factory default configurations. 5.5 Probe compensation Security When using the probe, in order to avoid electric shock, fingers should be kept behind the safety ring. Do not touch the metal part on the top of the probe when the probe is connected to the high voltage power supply. Before any measurement, connect the probe to the oscilloscope and ground the ground end.

Figure 5.1 Probe51 Manual probe compensation When the probe is connected to any input channel for the first time, perform this adjustment to match the probe to the input channel. Probes that are not compensated or compensated with bias can lead to measurement errors. To adjust the probe compensation, please follow the following steps: 1) Connect one end of the probe to the CH1 channel and the other end to the AFG channel. 2) Enter the oscilloscope interface, click Utility -> probe calibrate , and set the switch of the probe to 10X as prompted. Tablet1000 Quick guide Copyright © Qingdao Hantek Electronics Co., LTD 14