Quick start EN Figure 4.3 Probe41 Manual probe compensation When the probe is connected to any input channel for the first time, perform this adjustment to match the probe to the input channel. Probes that are not compensated or compensated with bias can lead to measurement errors. To adjust the probe compensation, please follow the following steps: 1) Connect one end of the probe to the CH1 channel and the other end to the AFG channel. 2) Enter the oscilloscope interface, click Utility -> Probe calibrate , and set the switch of the probe to 10X as prompted.
3) Check the shape of the waveform displayed Figure 4.4 Probe compensation4 3) If necessary, adjust the variable capacitance on the probe with a non-metallic screwdriver until the waveform displayed on the screen is "compensated correctly" as shown above. Tablet1000 User manua l Copyright © Qingdao Hantek Electronics Co., LTD 14