P648 Totalizer Reset, Internal Resets the internal totalizers. Entry: 0 = idle 1 = reset totalizer 2 2 = reset totalizers 1 and 2 End of Totalization parameters. Communication (P751 - P761) These parameters are specific to the use of the communication parameters. Refer also to Communications, page 47. P751 Baud Rate Sets the baud rate for the proprietary bi-polar current loop. This baud rate is not applicable to communication via Milltronics Comverter. Entry: 300, 1200, 2400, 4800 or 9600 baud 4800 baud is required for Dolphin communications over the bi-polar current loop. Parameters P760 Communication Mode Selects the communication mode. • bi-polar current loop:
interface with the host device (PLC or computer) • infrared link: communication is made using the Milltronics ComVerter. • maintenance: communication is made via the bi-polar current loop for RUN mode operations and infrared link for PROGRAM mode operations Entry: 1 = bipolar current loop 2 = infrared link 3 = maintenance P761 Identification Number Sets the identification number for the unit. Enter the desired identification number, range 0 - 15. End of communication parameters. 7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 61