Siemens — Test and Diagnostic (P900 - P951)

P941 Load Cell B, mV in Displays the mV signal input from the load cell. Range 0.00 - 60.00 mV. P942 V/F converter, V in Displays the input voltage to the voltage to frequency converter. Range 0 - 3.98 V P943 V/F converter, f out Displays the output frequency of the voltage to frequency converter. Range 0 - 131,072 P944 Power sensor Displays a voltage supply reference for diagnostic purposes. P949 Diagnostic Errors Enables or disables diagnostic error checking, E101 - E104 Entry: 0 = disable 1 = enable Parameters Refer to Troubleshooting on page 65. P950 Zero register Registers the number of zero calibrations that have been done since the last master reset.

P951 Span Register Registers the number of span calibrations that have been done since the last master reset. P952 Design Load Displays the value of the design load, which corresponds to the full scale value for alarm and mA output functions. The design load is calculated, based on the design rate and design speed. End of test and diagnostic parameters. 7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 63