:SEGMENTation:WRECord? /*The query returns 4096*/ :SEGMENTation:ENABle For example, SEGMENTation:ENABle ON /*Enable segment acquisition function*/ SEGMENTation:ENABle? /*The query returns 1*/ :HISTORY Command Subsystem The :HISTORY command is used to set the parameters related to waveform recording and playback. :HISTory:WRECord: For example, :HORTory:WRECord 16 /*Set the number of termination frames to 4096*/ :HORTory:WRECord? /*The query returns 4096*/ :HORTory:WRECord:FINTerval For example, :HORTory:WRECord:FINTerval 0.001 / * Set the time interval to 1ms * / :HORTory:WRECord:FINTerval?
/ * The query returns 1.000000e-03 * / 34 / 136 Command format :SEGMENTation:ENABle :SEGMENTation:ENABle? Function description Turn the segmented acquisition function on or off, or query the status of the segmented acquisition function. Parameter Name Type Range Default Bool {{1|ON}|{0|OFF}} 0|OFF Description Return format The query returns 1 or 0. Command format :HISTory:WRECord :HISTory:WRECord? Function description Set or query the historical waveform display positive sequence number. Parameter Name Type Range Default Integer 1 to the maximum number of frames that can be currently recorded 1 Description The maximum current recordable frame number can be queried by the: FUNCtion: WRECord:
FMAX? Command. Return format The query returns the current display frame number as an integer. Command format :HORTory:WRECord:FINTerval :HORTory:WRECord:FINTerval? Function description Set or query the time interval between frames during waveform playback. The default unit is s. Parameter Name Type Range Default Integer 10ms to 10s 10ms Return format The query returns the currently set time interval in scientific notation.