Hantek — :SEGMENTation:WRECord

For example, :DECoder1:POSition 300 /*Set the vertical position of the bus to 300*/ :DECoder1:POSition? /*The query returns 300*/ :TABLE:MODE For example, :TABLE:MODE SPI /*Set the decoder type to SPI*/ :TABLE:MODE? /*The query returns SPI*/ :TABLE:DISPlay For example, :TABLE:DISPlay ON /*Turn on the table decoder 1*/ :TABLE:DISPlay? /*The query returns 1*/ :SEGMENTation command subsystem The :SEGMENTation command is used to set the parameters for segmented acquisition. :SEGMENTation:WRECord For example, :SEGMENTation:WRECord 4096 /*Set the number of frames for segmented acquisition to 4096*/ 33 / 136 Return format The query returns an integer between 0 and 400.

Command format :TABLE:MODE :TABLE:MODE? Function description Sets or queries the type of the table decoder. Parameter Name Type Range Default Discrete {UART|LIN|CAN|SPI|IIC} UART Description UART, SPI, IIC, CAN, LIN correspond to UART decoding, SPI decoding and IIC decoding. LIN decoding, CAN decoding Return format The query returns UART, LIN, CAN, SPI or IIC. Command format :TABLE:DISPlay :TABLE:DISPlay? Function description Turn the form decoding function on or off, or query the form decoding function. Parameter Name Type Range Default Bool {{1|ON}|{0|OFF}} 0|OFF Description Only when the corresponding trigger is entered can the corresponding decoding be performed.

Return format The query returns 1 or 0. Command format :SEGMENTation:WRECord :SEGMENTation:WRECord? Function description Set or query the number of frames for segmented acquisition. Parameter Name Type Range Default Integer 80000 2 Description The number of frames can be set differently for different acquisition lengths Return format The query returns the number of segmented frames as an integer.