Hantek — :MASK:X

For example, :MASK:SOOutput ON /*Stop the output*/ :MASK:SOOutput? /*The query returns 1*/ :MASK:OUTPut For example, :MASK:OUTPut ON /*Turn on the sound prompt*/ :MASK:OUTPut? /*The query returns 1*/ :MASK:X For example, :MASK:X 0.28 /*Set the level adjustment parameter to 0.28div*/ :MASK:X? /*The query returns 2.800000e-01*/ :MASK:Y 42 / 136 Bool {{1|ON}|{0|OFF}} 0|OFF Description ▷On: When a failed waveform is detected, the oscilloscope will stop the test and enter the "STOP" state. At this time, the screen keeps displaying the measurement result (if the display is turned on), and the rear panel [Trigger Out] (if enabled) outputs only one pulse. ▷Off:

Even if a failed waveform is detected, the oscilloscope will continue the test, and the test results on the screen are continuously updated. The [Trigger Out] on the rear panel will output a pulse each time a failed waveform is detected. Return format The query returns 1 or 0. Command format :MASK:OUTPut :MASK:OUTPut? Function description Turn on or off the sound prompt when the test fails output, or query the status of the sound prompt. Parameter Name Type Range Default Bool {{1|ON}|{0|OFF}} 0|OFF Description ▷Off: When a failed waveform is detected, it will be displayed and output, but the buzzer will not alarm. ▷On:

When a failed waveform is detected, it will be displayed and output, and the buzzer will give an audible alarm (not related to the sound switch state) Return format The query returns 1 or 0. Command format :MASK:X :MASK:X? Function description Set or query the level adjustment parameter in the pass / fail test rule. The default unit is div. Parameter Name Type Range Default Integer 0.02 to 4 in steps of 0.02 0.02 Return format The query returns the current level adjustment parameter in scientific notation. Command format :MASK:Y :MASK:Y?