P350 Calibration Security This parameter provides additional security to the global lock (P000). Entry: 0 = view parameters, perform zero and span, no reset of totalizer 1 1 = same as level 0, but cannot perform span 2 = same as level 0, but cannot perform zero and span P359 Factoring Factoring is used as a method of calculating the value of the test load (P017) to a new physical test weight. Entry: 0 = idle 1 = factor Note: Totalization is halted during the factoring procedure, and resumed only upon return to the RUN mode. P360 Calibration Duration Parameters Sets the number of belt revolutions to use during a zero or span calibration. Enter number of belt revolutions, range 1 - 99.
P367 Direct Zero This parameter allows the zero reference count to be viewed or entered directly. Direct entry is intended for use when replacing software or hardware and it is not convenient to perform an initial zero at that time. P368 Direct Span Direct entry is intended for use when replacing software or hardware and it is not convenient to perform an initial span at that time. P370 Zero Limit This parameter allows the span reference count to be viewed or entered directly. Sets the zero calibration deviation limit from the last initial zero. If the accumulated deviation exceeds the limit, the zero calibration is aborted (E3). Entry:
0 = +/- 12.5% of initial zero 1 = +/- 2% of initial zero Page 56 Milltronics BW100 - INSTRUCTION MANUAL 7ML19985DJ01