Siemens — Troubleshooting

Troubleshooting Parameters 7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 65 message diagnosis action E1 (program) security code required enter access code into P000 E2 (run) totalizer resolution too low increase value (P631 - P639) E3 (run) zero out of range consider an initial zero P377, refer to Recalibration/Initial Zero on page 33 E4 (run) span out of range consider an initial span P388, refer to Recalibration/Initial Span on page 36 E5 (run) parameter not entered check parameters P005 - P017 for entry E6 (run) zero calibration required do a zero calibration E7 (run) span calibration required do a span calibration E8 (program) parameter value error check that value is valid E9(run) auto zero out of range auto zero has accumulated deviation beyond 2% from last operator initial zero.

If error is not caused by material on belt, then do an operator initiated zero. E10 (run) rate or span out of range loading on belt is 300% of rated load or greater. Investigate and if no mechanical cause, check to see if re-rating the design rate is required. E11 (run) speed greater than twice the design speed check design belt speed against actual belt speed, check speed constant, per- form speed constant adjust (P018) if necessary E12 (factoring) span out of range test load is either too low or beyond 100% of design load (P952). Replace test weight with a heavier or lighter weight and try factoring again.