For example, :TRIGger:WINDows:ALEVel 0.16 /*Set the trigger level upper limit to 160mV*/ :TRIGger:WINDows:ALEVel? /*The query returns 1.600000e-01*/ :TRIGger:WINDows:BLEVel For example, :TRIGger:WINDows:BLEVel 0.05 /*Set the trigger level lower limit to 50mV*/ :TRIGger:WINDows:BLEVel? /*The query returns 5.0000000-02*/ TRIGger:INTERVAl :TRIGger:INTERVAl:SOURce For example, :TRIGger:INTERVAl:SOURce CHANnel1 /*Set the trigger source to CH1*/ 78 / 136 Function description Set or query the upper limit of the trigger level when the super window is sent. The unit is the same as the current amplitude unit.
Parameter Name Type Range Default Integer (-5 × VerticalScale - OFFSet)至 (5 × VerticalScale - OFFSet) 2V Description This setting command is valid only when the selected source is an analog channel. Return format The query returns the upper trigger level in scientific notation. Command format :TRIGger:WINDows:BLEVel :TRIGger:WINDows:BLEVel? Function description Set or query the lower limit of the trigger level when the amplitude trigger is triggered. The unit is the same as the current amplitude unit.
Parameter Name Type Range Default Integer (-5 × VerticalScale - OFFSet)至 (5 × VerticalScale - OFFSet) 0 Description This setting command is valid only when the selected source is an analog channel. Return format The query returns the lower trigger level in scientific notation. Command format :TRIGger:INTERVAl:SOURce :TRIGger:INTERVAl:SOURce?
Function description Set or query the trigger source for interval trigger Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22|D23 |D30| D31|D32|D33|D40|D41|D42|D43| CHANnel1|CHANnel2|CHANnel3| CHANnel4} CHANnel1 Description Triggered when the interval between two consecutive rising edges (or falling edges) meets the set time condition ( , =,! =) Return format The query returns D10, D11, D12, D13, D20, D21, D22, D23, D30, D31, D32, D33, D40, D41, D42, D43, CHAN1, CHAN2, CHAN3, CHAN4.