:TRIGger:TIMeout:TIMe For example, :TRIGger:TIMeout:TIMe 0.002 /*Set the timeout time to 2ms*/ :TRIGger:TIMeout:TIMe? /*The query returns 2.000000e-03*/ TRIGger:TIMeout:LEVel For example :TRIGger:TIMeout:LEVel 0.16 /*Set trigger level 160mV*/ :TRIGger:TIMeout:LEVel /*The query returns 1.600000e-01*/ :TRIGger:WINDows :TRIGger:WINDows:SOURce For example, :TRIGger:WINDows:SOURce CHANnel2 /*Set trigger source to CH2*/ :TRIGger:WINDows:SOURce? /*The query returns CHAN2*/ :TRIGger:WINDows:ALEVel 77 / 136 Command format :TRIGger:TIMeout:TIMe :TRIGger:TIMeout:TIMe? Function description Sets or queries the timeout period triggered by the timeout. The default unit is s.
Parameter Name Type Range Default Integer 16ns to 10s 16ns Return format The query returns the timeout value in scientific notation. Command format :TRIGger:TIMeout:LEVel :TRIGger:TIMeout:LEVel? Function description : Set or query the trigger level when the timeout triggers Parameter Name Type Range Default Integer (-5 × VerticalScale - OFFSet)to (5 × VerticalScale - OFFSet) 0 Description This setting command is valid only when the selected source is an analog channel. Return format Returns the trigger level value in scientific notation Command format :TRIGger:WINDows:SOURce :TRIGger:WINDows:SOURce? Function description Set or query the trigger source triggered by the window.
Parameter Name Type Range Default Discrete {CHANnel1|CHANnel2|CHANnel3| CHANnel4} CHANnel1 Description Only analog channels can be used as trigger sources. Return format The query returns CHAN1, CHAN2, CHAN3, or CHAN4. Command format :TRIGger:WINDows:ALEVel :TRIGger:WINDows:ALEVel?