Hantek — :TRIGger:DELay:SB

For example, :TRIGger:RUNT:BLEVel 0.16 /*Set the trigger level lower limit to 160mV*/ :TRIGger:RUNT:BLEVel? /*The query returns 1.600000e-01*/ :TRIGger:DELay :TRIGger:DELay: SA For example, :TRIGger:DELay:SA CHANnel2 /*Set trigger source A to CH2*/ :TRIGger:DELay:SA? /*The query returns CHAN2*/ :TRIGger:DELay:SLOPA For example, :TRIGger:DELay:SLOPA NEGative /*Set the type of edge A to falling edge*/ :TRIGger:DELay:SLOPA? /*The query returns NEG*/ :TRIGger:DELay:SB 82 / 136 channel. Return format The query returns the lower trigger level in scientific notation. Command format :TRIGger:DELay:SA :TRIGger:DELay:SA?

Function description Set or query the trigger source of source A when delay trigger. Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22| D23|D30| D31|D32|D33|D40|D41|D42|D43| CHANnel1|CHANnel2|CHANnel3| CHANnel4} CHANnel1 Description When the digital channel is used as the trigger source channel, the digital probe must be connected to the oscilloscope. Return format The query returns D10, D11, D12, D13, D20, D21, D22, D23, D30, D31, D32, D33, D40, D41, D42, D43, CHAN1, CHAN2, CHAN3, CHAN4. Command format :TRIGger:DELay:SLOPA :TRIGger:DELay:SLOPA? Function description Sets or queries the edge type of edge A when the delay is triggered.

Parameter Name Type Range Default Discrete {POSitive|NEGative} POSitive Description : POSitive: rising edge trigger : NEGative: falling edge trigger Return format The query returns POS or NEG. Command format :TRIGger:DELay:SB :TRIGger:DELay:SB? Function description Set or query the trigger source of source B when the trigger is delayed. Parameter Name Type Range Default