For example, :TRIGger:LIN:ALEVel 0.16 /*Set trigger level 160mV*/ :TRIGger:LIN:ALEVel? /*The query returns 1.600000e-01*/ :TRIGger: CAN :TRIGger: CAN: SOURce For example, :TRIGger:CAN:SOURce CHANnel1 /*Set the trigger source to CH1*/ :TRIGger:CAN:SOURce? /*The query returns CHANnel1*/ :TRIGger:CAN:IDLe For example, :TRIGger:CAN:IDLe LOW /*Set idle low level*/ :TRIGger:CAN:IDLe? /*The query returns LOW*/ :TRIGger:CAN:BAUd 94 / 136 Return format Returns the trigger level value in scientific notation Command format :TRIGger:CAN:SOURce :TRIGger:CAN:SOURce? Function description Set or query the trigger source of CAN trigger.
Parameter Name Type Range Default Discrete {D10|D11|D12|D13|D20|D21|D22 |D23|D30|D31|D32|D33|D40|D41 |D42|D43|CHANnel1|CHANnel2| CHANnel3|CHANnel4} CHANnel1 Description When the digital channel is used as the trigger source channel, the digital probe must be connected to the oscilloscope. Return format The query returns D10, D11, D12, D13, D20, D21, D22, D23, D30, D31, D32, D33, D40, D41, D42, D43, CHAN1, CHAN2, CHAN3, CHAN4. Command format :TRIGger:CAN:IDLe :TRIGger:CAN:IDLe? Function description Set or query the idle level triggered by CAN Parameter Name Type Range Default Discrete HIGH Description HIGH: idle high LOW: idle low Return format The query returns LOW, HIGH Command format :
TRIGger:CAN:BAUd : TRIGger:CAN:BAUd? Function description Set or query the baud rate triggered by CAN. The default unit is bps. Parameter Name Type Range Default Discrete 10000