Hantek — :TRIGger:LIN:ALEVel

For example, :TRIGger:LIN:CONdition DATA /*set INTERVAL_END*/ :TRIGger:LIN:CONdition? /*The query returns INTERVAL_END*/ :TRIGger:LIN:ID For example, :TRIGger:LIN:ID 25 /*Set identifier to 25*/ :TRIGger:LIN:ID? /*The query returns 25*/ :TRIGger:LIN:DATA For example, :TRIGger:LIN:DATA 2 10 /*Set the data with index 2 to 10*/ :TRIGger:LIN:DATA? 2 /*The query returns 10*/ :TRIGger:LIN:ALEVel 93 / 136 End of ID field → LINID field data reception completion trigger. Sync code error → LIN sync field data reception is complete, but the data in the sync field is not equal to 0x55. Frame ID → LINID Field data reception is complete. Triggered when the ID data is equal to the ID set by the user.

Frame ID and data → LIN data is received normally, ID and data are both set by the user Return format The query returns INTERVAL FIELD | SYNC FIELD | ID FIELD | DATA | _ _ _ IDENTIFIER | ID DATA _ Command format :TRIGger:LIN:ID :TRIGger:LIN:ID? Function description Set or query LIN triggered identifier Parameter Name Type Range Default Discrete 0~26 −1 bit 00 Return format The query returns an integer Command format :TRIGger:LIN:DATA :TRIGger:LIN:DATA? Function description Set or query data values triggered by LIN. Parameter Name Type Range Default Discrete 0-3 Integer 0~255【XX】 XX Return format The query returns an integer Command format :TRIGger:LIN:ALEVel :TRIGger:LIN:ALEVel?

Function description : Set or query the trigger level during LIN trigger Parameter Name Type Range Default Integer (-5 × VerticalScale - OFFSet) to (5 × VerticalScale - OFFSet) 0 Description This setting command is valid only when the channel source of the data line is an analog channel.