Hantek — Appendix A: Technical Specifications

Appendix A: Technical Specifications All specifications herein mentioned apply to the series oscilloscopes. Before checking an oscilloscope to see if it complies with these specifications, make sure it meets the following conditions:  The oscilloscope must have been operating continuously for twenty minutes under the specified operating temperature.  The Do Self Cal operation must be performed through the Utility menu if the operating temperature changes by more than 5 ℃ .  The oscilloscope must be within the factory calibration interval. All specifications are guaranteed unless noted „typical‟. Oscilloscope Specifications Horizontal Vertical Note:

Bandwidth reduced to 6MHz when using a 1X probe. Acquisition Acquisition Modes Normal 31 Model Hantek2C72/Hantek2D72 Hantek2C42/Hantek2D42 Band Width 70MHz 40MHz Rising time ≤5ns ≤8.75ns Sample Rate Range 250MSa/s(Single-channel), 125MSa/s(Dual-channel) Waveform Interpolation (sin x)/x Record Length Max.

6K samples for single-channel, 3K samples per dual-channel SEC/DIV Range 5ns/div~500s/div 1, 2, 5 sequence A/D Converter 8-bit resolution, each channel sampled simultaneously VOLTS/DIV Range 10mV/div~10V/divat input BNC Measurement Range ±5div Selectable Analog Bandwidth Limit, typical 20MHz Low Frequency Response (-3db) ≤10Hz at BNC Rise Time at BNC, typical ≤5ns DC Gain Accuracy ±3% for Normal or Average acquisition mode, 10V/div to 10mV/div