Hantek — 5.3Channel configuration

Function introduction EN The display page is shown as following: Figure 5.2 Measurement bar chart 5.3 Channel configuration All settings on the page will be saved automatically at shutdown and loaded automatically at next startup. Press or to enter the or , and then press the configuration button to enter the page. use arrow buttons to select the configurations that you want to modify. Configurations descriptions: HTM200 user manual Copyright Qingdao Hantek Electronics Co., LTD 24 Configuration item Descriptions Type The thermocouple types include J/K/T/E/S/N/B/R/A/C/D. The initial is type K. Enable Open or close channels. Upper limit Set the upper temperature limit.

The initial value is 1800. When the temperature is higher than the upper limit, the color of the temperature values on the page will turn red, and a buzzer alarm will be sounded. The upper limit value can be set ranges from -200 ° C to 2480 ° C. Lower limit Set the temperature lower limit. The initial value is -200. When the