Load Cell Module DVP02LC-SL 4. Execute the reset to zero command by writing H’0001 into CR#26, as the WPLSoft program below. M0 5. Add a 1kg weight on the load cell. Please be aware of the maximum weight the load cell can take. 6. Write the weight of the weight (weight base point) into CR#33, as the WPLSoft program below. M2 7. Execute the weight base point command by writing H’0002 into CR#26, as the WPLSoft program below. M3 7.2 Correction by Software In this example, we will demonstrate how to correct CH1 on DVP02LC-SL by the software. 1. See 4.3 for how to wire the communication. Use a RS-232 communication wire to connect DVP02LC-SL to the PC and supply power to the two devices.

DVP-PLC Operation Manual 29 TO K100 K26 H1 K1 SET M0 Standard weight (1kg) Load cell DVP02LC-SL CH1 CH2 TO K100 K33 K1000 K1 SET M2 TO K100 K26 H2 K1 SET M3