Load Cell Module DVP02LC-SL 2. Connect the load cell to CH1. See 4.4 for external wiring. 3. Open the software and see 6.1 for how to set up the connection between the software and DVP02LC-SL. 4. Click “Parameters” on the left-hand side column to start setting up the parameters. Set up every parameter and eigenvalue according to the actual measuring requirements and specifications of the load cell. After the parameter setups are done, click “Download” to download them to DVP02LC-SL. 5. Click “Correction” on the left-hand side column to start correcting. First click “Reset to Zero Command” when there is nothing placed on the load cell. Now, the weight at CH1 is 0g. 6.
Add a 1kg weight on the load cell. Please be aware of the maximum weight the load cell can take. 7. Enter “1000” in “Zero Weight Value Correction” and click “Download” to download the settings to DVP02LC-SL. DVP-PLC Operation Manual 30 Load cell DVP02LC-SL CH1 CH2 Standard weight (1kg) Load cell DVP02LC-SL CH1 CH2