Siemens — Linearization Parameters (P390 - P396)

P377 Initial Zero The initial zero is the reference zero to which all subsequent operator initiated zero calibrations are compared in determining whether they have deviated beyond the zero limit (P370). Entry: 0 = idle 1 = initial zero Note: Refer to Recalibration/Initial Zero on page 33 for use of this function. P388 Initial Span The initial span is the reference to which all subsequent span calibrations are compared in determining whether they have deviated beyond 12.5% of the initial span. Entry: 0 = idle 1 = initial span Note: Refer to Recalibration/Initial Span on page 36 for use of this function. Parameters Linearization Parameters (P390 - P396) Note:

These parameters are used to compensate for non-linear response of the weighing system to the BW100. Refer to Recalibration/Linearization for details and example on the use of these parameters. P390 Linearization Enables or disables the linearization function. Entry: 0 = OFF 1 = ON P391 Linearizer, Point 1 Enter the load, in units of P017, for point 1 P392 Compensation, Point 1 Enter the calculated compensation, in percent, for compensation point 1 7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 57