8 Encoder diagnostics | Diagnosis for encoders with TTL/HTL/HTLs/ interface Deactivating the HSP function Certain encoders from HEIDENHAIN provide the HSP function that can be activated or deactivated. The HSP function is activated in the default setting. The HSP function should be deactivated only when the encoder is mounted. A corresponding message will then be shown in the PWT display screen. When an encoder is tested after it has been mounted, the HSP function should be activated. This default setting is activated when the PWT display screen is closed. For encoders with TTL interface, the HSP function is automatically set by the encoder when PWT switchover is activated.
Refer to the mounting instructions for the specific encoder. Tap More The More menu is opened Tap Functions The Display settings screen is opened Tap the Evaluation of incremental signals (HSP active) option The HSP function is deactivated Tap Back The changed settings take effect The Display settings screen is closed Inverting the reference pulse You can invert the evaluation of the encoder's reference pulse on the product. Tap More The More menu is opened Tap Functions The Function settings screen is opened Tap the Invert the reference pulse option The evaluation of the reference pulse is inverted.
Tap Back The changed settings take effect The Function settings screen is closed HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021 105