heidenhain PWT 100, PWT 101 — 8.1 Various diagnostics screens

Encoder diagnostics | Various diagnostics screens 8 8.1 Various diagnostics screens The product provides the following options for encoder diagnostics: If the Automatic Diagnosis menu is used, the product tries to automatically identify the encoder interface and run the appropriate diagnosis. Further information: "Performing Automatic Diagnosis", Page 55 The Manual Diagnosis menu enables you to manually set the encoder interface (for example, if the product does not automatically identify the encoder interface). Further information: "Performing Manual Diagnosis", Page 56 In both cases the device shows different views during the diagnosis, depending on the encoder interface.

The screens automatically adapt to the respective encoder function. The screens available in each case are shown in the following chart.

Automatic Diagnosis Manual Diagnosis Manual diagnostics settings Serial, serial + 1 Vpp, serial + TTL, serial + HTL*1 DRIVE-CLiQ TTL 1 Vpp, 11 µApp, 1 Vpp + Z1 HTL, HTLs Online diagnostics Level display Sine commutation (1 Vpp + Z1) PWT display Temperature Encoder information Encoder information Position 1/2 Position 2/2 PWT display (1 Vpp) Position Switching signals Counts Level display (TTL, HTL) PWT display ( *1 only via Manual Diagnosis ) If the Automatic Diagnosis does not identify an encoder correctly, you have to connect the encoder via the Manual Diagnosis menu. 54 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021