heidenhain PWT 100, PWT 101 — Diagnostic values of online diagnostics

Encoder diagnostics | Diagnosis for encoders with serial interface 8 Diagnostic values of online diagnostics Position display The Angle [degrees] (3) position display shows absolute or incremental position values depending on the encoder model. Abs : Absolute position value Inc : Incremental position value Inc shown in yellow: Reference mark not yet detected Tapping the display opens the Datum shift screen. Further information: "Datum shift screen ", Page 66 Inc shown in white: Reference mark detected Revolutions Revolutions (5) shows the number of revolutions performed depending on the encoder model. Tapping the display opens the Datum shift screen. Further information:

"Datum shift screen ", Page 66 Status The Status display (6) shows whether there are any messages about the connected encoder. Green: There are no messages Red: There are messages Tapping the display opens the Encoder status screen. Further information: "Encoder status screen", Page 70 Mounting clearance The values shown in Mounting clearance [mm] (4) make it easy to verify if the encoder is mounted correctly. Certain encoders generate values, such as the mounting clearance, that enable you to easily verify if the encoder is mounted correctly. If supported by the encoder, these values are transferred via the interface and displayed by the product in the Online diagnostics screen.

For the nominal values of the mounting clearance, refer to the mounting instructions for the respective encoder. 60 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021