heidenhain PWT 100, PWT 101 — 8.4.6 Level display screen

8 Encoder diagnostics | Diagnosis for encoders with serial interface 8.4.4 Position screen Figure 14: Position screen The Position screen displays information about a second position value. Precondition: The encoder is capable of displaying a second position value and supports this function (e.g. incremental encoders with purely serial interface or encoders that support functional safety). If the encoder also provides incremental signals, the second position value is generated from the incremental signals. When you call the Position screen, the second position is set to the starting value of the first position.

From that moment, the second position value is generated on the basis of the incremental signals. 8.4.5 PWT display screen If the encoder also provides 1 V PP incremental signals, you can evaluate them in the PWT display screen. The product automatically shows the PWT display screen. This enables you to check functions better. Further information: "PWT display screen ", Page 73 Information like, for example, the units for the count value display in Angle [degrees] or Position [µm] are read out and set automatically, depending on the encoder type. 8.4.6 Level display screen If the encoder also provides TTL incremental signals, you can evaluate them in the Level display screen.

The product automatically shows the Level display screen. This enables you to check functions better. Further information: "Level display screen for TTL/HTL/HTLs", Page 93 Information like, for example, the units for the count value display in Angle [degrees] or Position [µm] are read out and set automatically, depending on the encoder type. HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021 63