heidenhain PWT 100, PWT 101 — Resetting a datum shift

Encoder diagnostics | Diagnosis for encoders with serial interface 8 Datum shift for multiturn rotary encoders The following example applies to a multiturn rotary encoder and the setting [°] in the More menu. Tap More The More menu is opened Tap Datum Shift The Datum shift screen is opened Select or deselect the Datum shift compliant with incremental signals checkbox (where applicable) Further information:

"Assignment of zero position to signal period", Page 69 Tap Revolutions The product's virtual keyboard is opened Enter the value of the revolutions for the datum shift Tap OK Tap Position within one revolution [degrees] The product's virtual keyboard is opened Enter the value of the position within one revolution for the datum shift Tap OK Tap the Set button The datum is shifted The message Datum shift successful appears Tap OK If the datum shift was not successful, the product will display a corresponding message. Resetting a datum shift You can reset a datum shift you performed.

Tap More The More menu is opened Tap Datum Shift The Datum shift screen is opened Tap the Reset button The datum shift is reset The message Datum shift reset successful appears Tap OK If the datum shift was not successful, the product will display a corresponding message. 68 HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021