heidenhain PWT 100, PWT 101 — Display settings screen

8 Encoder diagnostics | Diagnostics for encoders with 1 V PP /11 µA PP /1 V PP + Z1 interface Clearing status messages Correct the error(s) displayed in the status message Tap Delete Status messages about errors that have been corrected will be deleted Status messages about errors that continue to exist will continue to be displayed If required, repeat this procedure until all pending status messages have been deleted Tap Back The screen that was last selected is opened Display settings screen Figure 35: Display settings screen The Display settings screen enables you to change the units of the measured values of the encoder. The setting has an effect on the count values displayed.

Selection Description Steps Incremental display of counting values. Indicated in signal periods. Rotatory The count values are displayed as Angle [degrees] . You have to enter the encoder's line count. Linear The count values are displayed as Position [µm] . You have to enter the signal period of the encoder. The choice of the count value unit affects all screens. Opening Display settings Tap More The More menu is opened Tap Units The Display settings screen is opened HEIDENHAIN | PWT 100, PWT 101 | Operating Instructions | 07/2021 89