US Army Materiel Command AMCP 706-117 — Environmental Series, Part Three: Induced Environmental Factors

Downloaded from https://www.everyspec.com AMCP 706.17 _ _ L +4 / +2 dD 0 - -+-2 -4 _______ (A) -in:Tcartridig - m X.W mm m m FUSem mWm I+4 / +2 0 - -2 --- _ __ .-- , - -4 7() 1-(. cartridge I 20 50 100 500 1,000 5,000 10k 20k FIGURE 7.3. Accuracy Curves of Condenser Microphone Output for Free. field a.5-in. and N-in. Carvldgn Mounted on a Tripod (solid lines sow nanrowband accuracy: daseftd line, broadband accuracy). a measure of a physical quantity, i.e., sound pressure. the curves relate to the entire system; i.e., microphone, This fact is important in evaluating the sound-level attenuator, amplifier, weighting networks, and indica- meter.

The sound-level meter differs from a voltmeter tor (meter). All three are also imprecise regarding the in that the frequency response of a voltmeter is made effects of extraneous factors such as teiperature, hu- as flat as possible, while that of the sound-level meter midity, and electromagnetic interference. In general, is deliberately altered by weighting networks to ac- the standards requite the manufacturer to state the count, to a first-order approximation, for the frequency range for such factors over which instrument accuracy response of the ear. Actually, three frequency-response is not adversely affected.

curves have been standardized because the response of At this point the standards diverge. The cur.es and the ear depends upon pressure level as well as fre- tolerances in JEC 123 specify performance in absolute qnency. These axe the so-called A, B, and C curves, and terms, i.e., with respect to a true sound-pressure level. approximate the inverse of the 40., 70-, and 100-phon The curves in IEC 179 and 81.4 are simple frequency- equal loudness curves of Fig. 7-1. response curves providing only rclative data.

In addi- The three most widely used standards describing tion, the A and B curves in SI,4 are relative to the C sound-level mesters are IEC Publication 123, Recom- curve, so the tolerances specified for the A and B curves mendetfon for Sound Levul Meters (Rof. 18); IEC Publi. must be added to those of the C curve. Table 7-5 lists caton 179, Preclsion Sound Weil Meter (Ref.. 19) and values and tolerances for F-11 three standards. (The tol. ANSI Standard 81.4, Spec/f cafon for Oenaml.?uqjae erances for the A and B curves under SI.4 are the total Sound Level Meters (Ref.

20).G A consideration of thes tolerances.) standards, Induding the similarities and differences, is The problem is to dcrive a set of figures from these relevant. data for absolute accuracy. Both IEC 179 and SI.4 allow AndCcurvesaall three statpeiy vrthaty the teanes fr B a ± I dB tolerance for absolute calibration at a refer- and C curves, and all three state that the tolerances for e c r q e c .A lt r esa d r sp o i etl r n e ence frequency. All three standards provide tolerances 6. A new ANSI standard (SI) became available In lae 1971. for making measurements at sound-pressure levels dif- It is similar to IEC 179 in many respeols but is more comprehen. ferent from the calibration level.

Included ire range 'iye. errors in jhe attenuator and meter (indicator), as well 7.13