Hantek — 4.3.4 Test Record Form

Performance verification Test EN 6. Once again, output a sinusoidal signal with an amplitude of 600 mVpp at other frequencies (not greater than 250MHz. The maximum frequency setting here varies depending on the model of the oscilloscope being tested) through Fluke5000. 7. Set the corresponding horizontal time base according to different frequencies. 8. Turn on the effective value measurement function, read and record the value (V1). 9. Calculate amplitude loss: Amplitude loss (dB)=20 Ч lg (V1/V) . 10. Keep the other settings of the oscilloscope in step 3 unchanged and set the vertical gear to 200 mV/div. 11.

Output a sine signal with a frequency of 1 MHz and an amplitude of 1.2 Vpp through Fluke5000. 12. Repeat steps 5-9. 13. Keep the other settings of the oscilloscope in step 3 unchanged and set the vertical gear to 2V/div. 14. Output a sine signal with a frequency of 1 MHz and an amplitude of 5.5 Vpp through Fluke5000. 15. Repeat steps 5-9. 16. Close CH1. Test CH2, CH3, and CH4 in sequence according to the above testing method and record the test results.

4.3.4 Test Record Form DPO7000 Performance Verification Guide Copyright © Qingdao Hantek Electronics Co., LTD 14 Voltage Fluke Amp Fluke Freq Meas Range Passed Range Result or not 100mV/div 600mV 1MHz -3dB to 3dB 10MHz 20MHz 50MHz 100MHz 200MHz 250MHz