Preparation before use EN ⚫ Connect the grounding crocodile clamp wire of the probe to the grounding terminal in the figure below. ⚫ Connect the probe to the input terminal of channel 1 and the compensation signal output terminal marked in the figure below. ⚫ Set the probe attenuation ratio to X10 and click on the button area [Auto Scale] . ⚫ Observe the pictures on the waveform to see if the square wave signal is displayed normally. As shown in the following figure. ⚫ Check CH2~CH4 channels using the same method. If the square wave waveform displayed in the time base does not match the figure above, you can follow the instructions in the section on Probe Compensation . Warning:
To avoid electric shock when using the probe, please make sure that the insulated wire of the probe is in good condition. Do not touch the metallic part of the probe when the probe is connected to high voltage source. Tip: The probe compensation signal can only be used for probe compensation DPO7000 Quick Guide Copyright © Qingdao Hantek Electronic Co., LTD 16