Display EN 12.1 To Select the Display Type Click on the Type menu label, which can be selected as vector or point. Choose whether the waveform is displayed as a vector or as a point. The vector setting will fill in the blank space between adjacent sampling points in the display; Point settings only display sampling points. The vector type sampling points are displayed through connecting lines. This mode provides the most realistic waveform in most cases, making it easy to view the steep edges of waveforms (such as square waves).
12.2 To Set the Persistence Time In the displayed menu, click on the Persis.Time menu label and select afterglow time (Min, 100ms, 200ms, 500ms, 1s, 2s, 5s, 10s, Infinite). ⚫ Minimum value: Observable waveform with high refresh rate changes. ⚫ Specific value: burrs with slower changes or lower probability of occurrence can be observed. The afterglow time can be set to 100ms, 200ms, 500ms, 1s, 2s, 5s, and 10s. ⚫ Infinite: When the oscilloscope displays the newly collected waveform, it will not clear the previously collected waveform. The collected waveforms will be displayed in colors with lower brightness, while the newly collected waveforms will be displayed in normal brightness and color.
The use of infinite afterglow can measure noise and jitter, capturing occasional events. 12.3 To Set the Waveform Intensity Click on the Intensity menu button, and set the brightness value directly through the multi-functional knob V0 or the numeric keyboard, with a range of 0%~100%. 12.4 To Set the Screen Grid Click on the Grid menu tab to select the types of screen grids: Dot, Line, and Close. ⚫ Dot: Set the background grid line as a point. ⚫ Line: Set the background grid to solid lines. ⚫ Close: Close the background grid. DPO7000 User manua l Copyright © Qingdao Hantek Electronic Co., LTD 128