System Utility Function Setting EN settings. In the Save Group option, the available storage options are "Save Image", "Save Wave", or "Save Setup". After setting up, click the [Quick Action] button to store the selected content according to the configuration. Do not insert a USB drive, it is saved internally by default. Insert a USB drive and save it in the USB drive. 13.10 Screen Saver Click [Utility] -> More -> Screen Saver menu tab, options: off, Picture, Text. Select Picture: Click to Select Picture and load it locally on the machine or on a USB drive. The image format is *. png, *. bmp, *. jpg, *.

tif; Click on the Time to Start menu tab and set the waiting time (5min, 10min, 30min, 1h, 2h, 4h); Click on the Preview menu tab to see the effect of the settings in advance; Click on the Default Menu tab to restore the settings to their default state. Select text: Click on the Text menu label and enter custom text on the keyboard; Click on the Time to Start menu tab and set the waiting time (5min, 10min, 30min, 1h, 2h, 4h); Click on the Preview menu tab to see the effect of the settings in advance; Click on the Default Menu tab to restore the settings to their default state. 13.11 Self Check Enter the self-test menu through the following methods:

⚫ Click on the [Utility] -> More -> More -> Self Check menu tab to enter the Self Check menu. ⚫ Click on the navigation icon [Hantek] in the upper left corner of the screen, select Self Check , and enter the Self Check menu. The self-test options include Key Test, Screen Test, and Touch Test. Key Test: Click on the Key Test menu tab to enter the keyboard testing interface. The icon corresponding to the button appears on the screen. By pressing the button on the front panel of the instrument or rotating the knob, the corresponding icon can be illuminated to determine whether the button or knob function is normal. Press' Run/Stop 'three consecutive times to exit the testing function.

Screen Test: Click on the Screen Test menu tab to enter the screen testing interface. Red and XXX appear in sequence. Follow the prompts and press "Single" to enter the next screen. Press' Run/Stop 'three consecutive times to exit the testing function. Touch Test: Click on the Touch Test menu tab to enter the touch function testing interface. Use your finger to click on the screen. If there are corresponding lines displayed in the blank area and the boxes that pass through turn into green background boxes, it indicates that the touch screen function is normal in that area. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 145