Hantek — DPO7000 UserManual

To Set the Sample System EN 7.7 Sample Rate Sampling refers to the process in which an oscilloscope converts analog signals into digital signals at certain time intervals and stores them in sequence. The sampling rate is the reciprocal of the time interval. Open [Acquire] and the sampling rate menu will display the current sampling rate value. The sampling rate is displayed in the status bar at the top of the screen, and can be indirectly changed by adjusting the horizontal time base or modifying the storage depth. The impact of low sampling rate on waveform: ⚫ Waveform distortion:

Due to the low sampling rate, some waveform details are missing, resulting in significant differences between the waveform displayed by the oscilloscope sampling and the actual signal. ⚫ Waveform aliasing: since the sampling rate is lower than 2 times of the actual signal frequency (Nyquist frequency), the waveform frequency when reconstructing the sampled data is lower than the frequency of the actual signal. ⚫ Waveform leakage: Due to the low sampling rate, the waveform reconstructed from the sampled data did not reflect all the actual signals. Copyright © Qingdao Hantek Electronic Co., LTD DPO7000 User manual 55