Hantek — 3.7.1 SOURce<n>:DATA:DAC16:BIN

Command system EN Instructions None Return format Return the current gate time. Example :COUNter:GATEtime 0.01/* set the gate time of the frequency meter to 10 ms*/ :COUNter:GATEtime?/* Query returns 10ms*/ 3.6.3 COUNter:MEASure Command format COUNter: MEASure? Function description Query the current measurement results of the frequency meter. Argument None Instructions None Return format Return the measurement results of each argument in the form of "frequency, period, duty period, positive pulse width, negative pulse width", where each argument is represented as a scientific count. Example COUNter:MEASure?

/* Query returns 1.000099993E+03,9.999000134E-04,1.422600068E+01,1.422537019E-04,8.57646311 5E-04*/ 3.7 DATA subsystem 3.7.1 SOURce :DATA:DAC16:BIN Command format Copyright © Qingdao Hantek Electronics Co., LTD HDG300C Programming manual 23